Digital Oceanographic Thermometer 
SBE 38

[HRule Image]

Brochure/specifications in pdf format     Manual     Reference Sheet     Configuration

SBE 38DESCRIPTION

Sophisticated A/D acquisition electronics, ultra-stable thermistor, and state-of-the-art calibration provide the standards-level performance of an expensive AC bridge and platinum thermometer at a small fraction of the cost. The SBE 38 is unaffected by shock and vibration, has high accuracy and stability, and is easy to use. It has a rugged corrosion-proof 10,500 meter (34,400 ft) titanium pressure housing. Real-time temperature data is transmitted in ASCII characters (in °C or raw counts) via an RS-232 or optional RS-485 serial interface for display or logging by PC or data logger.

The SBE 38's measurement range is -5 to +35 °C; absolute accuracy is better than 0.001 °C (1 mK) and resolution is approximately 0.00025 °C (0.25 mK). Each sensor includes certification that demonstrates drift of less than 0.001 °C during a six-month period.

Applications include calibration baths, oceanographic/aquatic research, and environmental monitoring.

OPERATION

The SBE 38 operates in one of three modes:

On power up, the SBE 38 reads its EEPROM, which includes calibration coefficients and other setup information. As programmed, the SBE 38 samples and transmits temperature continuously, or waits for a command. For RS-485 with several sensors sharing one pair of wires, the SBE 38 cannot sample continuously.

MEASUREMENT METHOD

Temperature is determined by applying an AC excitation to reference resistances and an ultra-stable aged thermistor with a drift rate of less than 0.002 °C per year. Each of the resulting outputs is digitized by a 24-bit A/D converter. The reference resistor is a hermetically sealed VISHAY. AC excitation and ratiometric comparison using a common processing channel removes measurement errors due to parasitic thermocouples, offset voltages, leakage currents, and gain errors. The maximum power dissipated in the thermistor is 0.5 microwatts, and contributes less than 200 µK of overheat error.

A raw count (ratio) is related to resistance measurements:

Raw Counts = 1048576 * NT / NR
where: NR is the output from the reference resistor and NT is the thermistor output.

The number of acquisition cycles (raw counts) averaged per sample is user-programmable. Increasing the number of cycles per sample increases the time to acquire the sample and the interval between samples, while reducing the RMS temperature noise from the sensor. The interval between samples is:

interval [seconds] = (0.133 * Navg) + 0.339
where: Navg is the number of acquisition cycles per sample.

The SBE 38’s converted output is computed from the raw count and calibration coefficients that are stored in EEPROM.

CALIBRATION

The SBE 38 is calibrated in Sea-Bird's state-of-the-art calibration laboratory, which maintains primary temperature standards (water triple point [TPW] and gallium melting point [GaMP] cells), ITS-90 certified and standards-grade platinum resistance thermometers, and a low-gradient temperature bath.

Temperature is computed using the Steinhart-Hart polynomial for thermistors (Steinhart and Hart, 1968; Bennett, 1972), which is based on thermistor physics. The equation characterizes the non-linear temperature vs. resistance response of the sensor. Note that thermistors require individualized coefficients to the Steinhart-Hart equation, because the thermistor material is an individualized mix of dopants:

t90L = { [ 1.0 / (a0 + a1ln(n) + a2ln²(n) + a3ln³(n) ) ] - 273.15}* Slope + Offset       [°C]

where: n is the SBE 38 output.

SPECIFICATIONS

Interface:

Housing: Titanium, rated at 10,500 meters (34,400 feet)

Weight:

Range: -5 to +35 °C

Initial Accuracy 1: ± 0.001 °C (1 mK)

Resolution: 0.00025 °C (0.25 mK)

Stability: 0.001 °C (1 mK) in six months, certified

Response Time 2: 500 milliseconds

Self-heating Error: less than 200 µK

1NIST-traceable calibration applying over the entire range.
2Time to reach 63% of final value following a step change in temperature.

 

Actual Calibration Data for Sensor Number 80
Calibration Date: 02 September 1997
a0 = -2.809379e-05   a2 = -2.619655e-06
a1 = 2.783484e-04   a3 = 1.598734e-07

Bath 
Temperature 
(°C)
Instrument 
Output 
(n)
Instrument 
Temperature 
(°C)
Residual 
(Instrument - Bath)
(°C)

-1.52985

824162.7

-1.52983

0.00002

1.03108

733633.1

1.03106

-0.00002

4.60520

625547.1

4.60518

-0.00002

8.11169

536776.4

8.11169

-0.00000

11.61533

462132.6

11.61536

0.00003

15.17575

398167.3

15.17574

-0.00001

18.63931

345476.6

18.63934

0.00003

22.14032

300170.8

22.14031

-0.00001

25.66793

261276.6

25.66793

0.00000

29.13948

228549.1

29.13944

-0.00004

32.61481 200420.3 32.61484 0.00003

Click on the thumbnail to view sample calibration graph.

 38calgraph.jpg (54732 bytes)

ADDITIONAL INFORMATION / LINKS: