APPLICATION NOTE NO. 34

Instructions for use of Conductivity Cell
Filling and Storage Device P/N 50087 and 50087.1

Revised January 2005

Application note in pdf format

This application note provides instructions for use of PN 50087 / 50087.1 syringe and tubing assembly in rinsing, cleaning, and storing conductivity sensors. The tubing assembly consists of a length of 6.35 mm (1/4 inch) I.D. tube connected by a plastic reducing union to a short piece of 11.1 mm (7/16 inch) I.D. tube. Refer to Application Note 2D: Instructions for Care and Cleaning of Conductivity Cells for information on water and solutions recommended for use.

Note: The procedure below can also be used with the SBE 16 and SBE 19, which are no longer in production.

Procedure for Use

  1. To fill the conductivity cell, draw about 40-60 cc of purified water into the syringe.
  2. Connect the plastic tubing to the TC duct intake on the temperature sensor [Figure 1] (or to the open end of the conductivity cell on systems without the TC duct [Figure 2]), and inject solution into the cell and pump plumbing.
  1. (If applicable) Loop the rubber band around a bar on the CTD cage and back over the top of the syringe to secure the apparatus for storage.

REMOVE THE SYRINGE AND TUBING ASSEMBLY BEFORE DEPLOYMENT !