APPLICATION NOTE NO. 64-1

Plumbing Installation -- SBE 43 DO Sensor and Pump on a CTD
September 2003

div.gif (3429 bytes)

Application note in pdf format

This Application Note describes installation of an SBE 43 Dissolved Oxygen (DO) Sensor and pump (and associated plumbing) on an SBE 9plus CTD, using PN 90087 Universal Plumbing Kit. The plumbing guidelines for the SBE 9plus also apply to an SBE 19, 19plus, or 25 CTD, except as noted.

Preparation

Parts Included in PN 90087 Universal Plumbing Kit

PN Description Quantity
23337 Pump air release valve -- for vertical configuration only 1
30387 Tube, 1/2" Y-fitting -- for connection to pump air release valve, vertical configuration only 1
30388 Tygon tubing, 19 mm (0.75 in.) OD, 13 mm (0.5 in.) ID 4 ft
30579 Tygon tubing, 13 mm (0.5 in.) OD, 9.5 mm (0.375 in.) ID -- for connections to DO sensor 0.5 ft
30389 Cable tie, 102 mm (4 in.) -- to secure tubing connections 20
30458 Cable tie, 381 mm (15 in.) -- to secure Y-fitting to main housing 2
     

Tools Required

Sharp knife or scalpel for cutting tubing
Straight-edge ruler for measuring and cutting

Installation

Note for SBE 9plus or 25:
If your conductivity sensor does not have a conductivity tube support, install the conductivity tube support kit before installing the plumbing. (See Application Note 36 for installation details)

Conductivity tube support kit:
Part Number 50094 for aluminum sensors
Part Number 50108 for titanium sensors

Basic considerations for plumbing installation:

Plumbing installation details differ, depending on whether the sensors will be deployed in a horizontal or vertical configuration. See Horizontal Configuration Schematic and Details and Vertical Configuration Schematic and Details below.

 

Horizontal Configuration Schematic and Details

The horizontal configuration is most commonly used when the CTD is integrated with an SBE 32 Carousel Water Sampler with an extension stand. When installing plumbing for the horizontal configuration:

Failure to configure the system in this manner can trap air, preventing the pump from working properly.

If the system does not include a DO sensor, connect the tubing from the conductivity cell directly to the pump intake.

DO sensor intake and exhaust tubing detail:

Secure each tubing connection with two cable ties (not shown).

 

 

 

Vertical Configuration Schematic and Details

When installing plumbing for the vertical configuration:

DO sensor intake and exhaust tubing detail:

Secure each tubing connection with two cable ties.

Note: Exhaust tubing is usually omitted by Sea-Bird for an SBE 19, 19plus, or 25.