APPLICATION NOTE NO. 72
Setting Up WET Labs ECO-FL Fluorometer with
Bio-Wiper™
May 2005
![]()
Application note in pdf format
This application note applies to the following models of the WET Labs ECO-FL Fluorometer with Bio-Wiper:
ECO-FL-NTUS chlorophyll and turbidity sensor
ECO-FLS chlorophyll sensor
These sensors can be configured to operate in profiling mode or in moored mode:
Profiling mode – The sensor opens the bio-wiper when power is applied and keeps it open continuously, even when power is removed. The fluorometer samples continuously when powered, and outputs a voltage that changes as the data changes.
Moored mode – The sensor opens the bio-wiper when power is applied, takes a sample, outputs a voltage, closes the bio-wiper, and holds the voltage until power is removed.
When Sea-Bird integrates the sensor with one of our CTDs, we set up the sensor to operate in the mode appropriate for the CTD. However, you may want to change the mode in the future, if:
To change the ECO-FL configuration, connect the ECO-FL directly to the computer and to a 12 V power source, using the cable supplied by WET Labs. Configure the ECO-FL using a terminal program (WET Labs provides a graphical interface program, ECOView, which can be used instead of a terminal program – see the WET Labs website [www.wetlabs.com] for details).
The required comm port settings are: 19200 baud, 8 data bits, no parity.
| Configure ECO-FL for Profiling Mode | Configure ECO-FL for Moored Mode |
| Send the following commands: !!!!! $set 0 $pkt 0 $sto |
Send the following commands: !!!!! $set 1 $pkt 1 $sto |
|
Notes:
|
|
After you have configured the ECO-FL for your CTD, reconnect the ECO-FL to the CTD.
General
Delay before sampling sets the amount of time to wait after switching on external voltages before sampling. The ECO-FL requires approximately 3.5 seconds to sample, if set up to take a single measurement for each sample (longer time is required for multiple measurements per sample). Therefore, set:
- DelayBeforeSampling= (16plus RS-232 version or 19plus in moored mode), or
- #iiDelayBeforeSampling= (16plus RS-485 version or 16plus-IM)
to greater than the expected time required for the ECO-FL to sample.
Older CTD firmware
SBE 16plus (RS-232) firmware version < 1.6c,
SBE 16plus (RS-485) Digital firmware version < 1.0a,
SBE 16plus-IM Digital firmware version < 1.1,
SBE 19plus firmware version < 1.4dIn older CTD firmware versions, power was not provided for enough time for the Bio-Wiper to close after the status command (DS for 16plus RS-232 version and 19plus; #iiDS for 16plus RS-485 version and 16plus-IM) was sent if the Bio-Wiper was in Moored mode. If the user did not notice that the Bio-Wiper was still open, and then deployed the CTD on a mooring with a delayed start time, the ECO-FL could become fouled, because the Bio-Wiper would not close until after the first sample was taken.
For these older firmware versions, after sending the status command:
Send the TV, TVR, or TS command (#iiTV, #iiTVR, or #iiTS for 16plus RS-485 version and 16plus-IM) and then wait at least 10 seconds to cycle the Bio-Wiper open/closed. Verify that the Bio-Wiper has closed before deploying.
Newer CTD firmware
SBE 16plus (RS-232) firmware version > 1.6d (October 2004 and later)
SBE 16plus (RS-485) Digital firmware version > 1.0c (May 2005 and later),
SBE 16plus-IM Digital firmware version > 1.1b (May 2005 and later)
SBE 19plus firmware version > 1.4f (May 2005 and later)Sea-Bird added a Bio-Wiper command to the command set, to ensure that the Bio-Wiper automatically closes after sending the status command:
Command Description Biowiper=x
(16plus RS-232 version and 19plus)or
#iiBiowiper=x (16plus RS-485 version and 16plus-IM)
x=Y: Configuration includes WET Labs ECO-FL fluorometer with Bio-Wiper. CTD is powered longer (a total of 8 seconds) for status command, providing 4 seconds for Bio-Wiper to open and then shut again if in Moored mode.
Note: 4 seconds for the Bio-Wiper to open and shut is sufficient for the typical application, with the ECO-FL set up to take a single measurement for each sample. However, the ECO-FL can be user-programmed to take and average a number of measurements for each sample; if averaging multiple measurements, 4 seconds may not provide sufficient time for the Bio-Wiper to close. Use the TV, TVR, or TS command (#iiTV, #iiTVR, or #iiTS for 16plus RS-485 version and 16plus-IM) after the status command, as described above for older CTD firmware, to ensure the Bio-Wiper closes for these applications.
x=N (default): Configuration does not include ECO-FL with Bio-Wiper.
![]()