APPLICATION NOTE NO. 09
Application note in pdf format
Some Sea-Bird CTDs can be equipped with a Sea Tech Fluorometer or WET Labs Flash Lamp Fluorometer. For these fluorometers, the CTD configuration file requires the scale factor and offset to enable the software to calculate the concentration:
Concentration = (voltage * scale factor / 5) + offset
Sea Tech Fluorometers and WET Labs FLFs have internal, switch-selectable ranges. The scale factor is dependent on the fluorometer range:
Fluorometer | Range (milligrams/cubic meter) | Scale Factor |
Sea Tech | 0 - 3 | 3 |
0 - 10 (factory-set default) | 10 | |
0 - 30 | 30 | |
0 - 100 | 100 | |
0 - 300 | 300 | |
0 - 1000 | 1000 | |
WET Labs FLF | 0 - 100 | 100 |
0 - 300 (factory-set default) | 300 | |
0 - 1000 | 1000 | |
The offset can be calculated by measuring the fluorometer voltage output when its light sensor is completely blocked from the strobe light by an opaque substance such as heavy black rubber:
offset = (-scale factor * output voltage ) / 5
NOTE: The scale factor and offset can be adjusted to fit a linear regression of fluorometer responses to known chlorophyll a concentrations.
NOTE: Current versions of Sea-Bird Windows software list the Seatech/WET Labs FLF under Fluorometers when modifying the .con file.
- If the WET Labs FLF is not listed in the software version you are using, select the Sea Tech Fluorometer and enter the correct scale factor for the WET Labs FLF from the table above.
- Display Setup appears. Click on Modify Display Parameters.
- Fixed Display Variables Setup appears. In the desired row, click Select Variables.
- Select a Variable from the List appears. Double click on the desired voltage from the list of available voltage numbers (see CTD Configuration Sheet at beginning of CTD manual for voltage number).
- Fixed Display Variables Setup appears again. Click OK.
- Display Setup appears again. Click OK.
- You May Have Changed Your Display Parameters appears. Click Yes for Save Display Parameters to Disk.
- In the Save dialog box, select the desired location and file name, and click OK.
- Verify that the COMM port is correct.
- For SBE 16 / 16plus only -- Start CTD logging.
- Click Start Acquire. (Note: SBE 9plus will start logging automatically.)
- For SBE 19 / 19plus and 25 only -- Start CTD logging.