APPLICATION NOTE NO. 11
QSP-L
Calculating Calibration Coefficients for Biospherical PAR Light Sensor
with Built-In Log Amplifier
Revised June 2005
Application
note in pdf format
This application note applies to the following
Biospherical Instruments PAR
light sensors, which all have a built-in log amplifier:
- QSP-200L and QCP-200L -- no longer in production
- QSP-2300L, QCP-2300L, and MCP-2300 -- current production
These PAR sensors are compatible with the following Sea-Bird CTDs:
- SBE 9plus
- SBE 16 or 19 – These PAR sensors may not be compatible with 6-cell
housing version of these CTDs; consult Sea-Bird.
- SBE 16plus,
16plus-IM, or
19plus – CTD’s
optional PAR connector not required when using one of these PAR sensors. The PAR sensor interfaces with an A/D voltage channel
on the CTD.
- SBE 25 – CTD’s PAR
connector (standard on current production SBE 25s, optional on older versions)
not used with these PAR sensors. The PAR sensor interfaces with an A/D voltage
channel on the CTD.
Note: The CTD voltage channel for use with the PAR sensor can be
single-ended or differential.
SEASOFT computes PAR
using the following equation:
PAR = [multiplier * (109 *
10(V-B)
/ M) / calibration constant] + offset
Enter the following coefficients in the CTD configuration file:
M = 1.0 (Notes 2 and 3)
B = 0.0 (Notes 2 and 3)
calibration constant
= 105 / Cw (Notes 2 and
4)
multiplier = 1.0 for output units of
μEinsteins/m2.sec (Note 5)
offset =
- (104 * Cw * 10V) (Note 6)
Notes:
- Edit the CTD configuration (.con) file using the Configure menu (in SEASAVE
or SBE Data Processing
in our SEASOFT-Win32 suite of programs)
or SEACON (in SEASOFT-DOS).
- Sea-Bird provides two calibration sheets for the PAR sensor in the CTD
manual:
- Calibration sheet generated by Biospherical, which contains
Biospherical’s calibration data.
- Calibration sheet generated by Sea-Bird, which incorporates the
Biospherical data and generates M, B, and calibration constant
needed for entry in Sea-Bird software (saving the user from doing the math).
- For all SBE 911plus, 16, 16plus, 16plus-IM, 19, 19plus,
and 25 CTDs, M = 1.0. For SBE 9/11 systems
built before 1993 that have differential input amplifiers, M = 2; consult your SBE 9 manual or contact factory for further information. B should always be set to 0.0.
- Cw is the wet
μEinsteins/cm2.sec/"amps"
coefficient from the Biospherical calibration sheet. A typical value is
4.00 x 10 -5.
- The multiplier can be used to calculate irradiance in units other than
μEinsteins/m2.sec.
See Application Note 11General for multiplier
values for other units.
The multiplier can also be used to scale the data, to compare the shape of data sets taken at disparate light levels. For example, a multiplier of 10
would make a 10
μEinsteins/m2.sec light level plot as 100
μEinsteins/m2.sec.
- Offset (μEinsteins/m2.sec) = -
(104 * Cw * 10V), where
V is the dark voltage.
For typical values (Cw =
4.00 x 10 -5 and Dark Voltage = 0.150), offset =
-0.5650.
The dark voltage may be obtained from:
- Biospherical calibration certificate for your sensor, or
- CTD PAR voltage channel with the sensor covered (dark) -- in SEASAVE,
display the voltage output of the PAR sensor channel.
Instead of using the dark voltage to calculate the offset, you can also
directly obtain the offset using the following method: Enter M, B, and Calibration
constant, and set offset = 0.0 in the .con
file. In SEASAVE, display the calculated PAR output with the sensor
dark; then enter the negative of this reading as the offset in the .con file.
Mathematical Derivation
- Using the sensor output in volts (V), Biospherical calculates:
light (μEinsteins/cm2.sec) = Cw * (10Light Signal Voltage - 10Dark
Voltage).
- SEASOFT calculates
μEinsteins/m2.sec = [multiplier
*
109 * 10(V - B) / M) / Calibration constant] + offset
where M, B, Calibration constant, and offset are the SEASOFT coefficients entered in the CTD
configuration file.
- To determine Calibration constant, let B = 0.0, M = 1.0, multiplier = 1.0. Equating the
Biospherical and SEASOFT relationships:
104 (cm2/m2)* Cw * (10Light Signal Voltage
- 10Dark Voltage) = (109 * 10V) / Calibration
constant +
offset
Since offset = - (104 * Cw * 10Dark
Voltage), and V = Light Signal Voltage:
Calibration constant = 109 / (104 * Cw) = 105 / Cw
Example: If Wet calibration factor = 4.00 * 10-5
μEinsteins/cm2.sec,
then C = 2,500,000,000 (for entry into .con file).
Notes:
- See Application Note 11S for integrating a Surface
PAR sensor with the SBE 11plus
Deck Unit (used with the SBE 9plus CTD).
- See Application Note 47 for integrating a Surface
PAR sensor with the SBE 33 or
36 Deck Unit (used with the
SBE 16, 16plus, 19, 19plus, or 25 CTD).
Last modified:
06-Apr-2007
Sea-Bird Home Phone: 425-643-9866
Fax: 425-643-9954 E-mail:
seabird@seabird.com