APPLICATION NOTE NO. 62

Calculating Calibration Coefficients for the WET Labs ECO-AFL or ECO-FL Fluorometer, or ECO-FL-NTU Fluorometer/Turbidity Meter
December 2006
div.gif (3429 bytes)

Application note in pdf format

This Application Note applies to the following WET Labs fluorometers:

 

Fluorometer Calibration Coefficients

The fluorometer supplied with your Sea-Bird CTD has a response that is linear over the measurement range provided.

Nominal Range (µg/l) 0.01 to 30 0.01 to 60 0.01 to 120
Analog Scaling Value / Range (enter in ECOView Host) 1 2 (factory default) 4
       

Select the ECO-AFL/FL fluorometer when setting up the configuration (.con) file in Sea-Bird software (SBE Data Processing or SEASAVE in SEASOFT-Win32 or SEACON in SEASOFT-DOS). The software prompts for Vblank and Scale Factor and calculates chlorophyll concentration as:

concentration (µg/l) = (Vsample - Vblank) * Scale Factor     (see Note)

where:

Vsample (volts) = in-situ output of the fluorometer

Vblank (volts) = measured output for a sea water blank (pure, de-ionized water)     (see Note)

Scale factor (µg/l-volts) = multiplier

The fluorometer comes with a calibration sheet that lists values for Vblank and Scale Factor (see Note). If you changed the analog scaling value (ECO-FL series only), change the Scale Factor to correspond.

Note: Calibration sheets from WET Labs for newer fluorometers may list Dark Counts instead of Vblank. Use the Dark Counts value in place of Vblank when setting up the configuration (.con) file.

Example Concentration Calculation in Sea-Bird Software:

Vblank = 0.05 volts and Scale Factor = 12.35 µg/l-volts (from calibration sheet)

Measured voltage from fluorometer = Vsample = 4.65 volts

Calculated concentration (µg/l) = (Vsample - Vblank) * Scale Factor
= (4.65 - 0.05) * 12.35 = 56.8 µg/l

While the factory-supplied Scale Factor can be used to obtain approximate values, field calibration is highly recommended. The relationship between fluorescence and chlorophyll-a is highly variable, and is not easy to determine in the laboratory. Species distribution, ambient light level, and health of the stock are just some of the factors that affect the relationship. To accurately measure chlorophyll-a concentration with a fluorometer, perform calibrations on seawater samples with concentrations of plankton populations that are similar to what is expected in-situ. Determine chlorophyll-a concentrations independently, and use those concentrations, as well as readings from the fluorometer, to determine the correct Scale Factor. It is only through the use of these calibrations that a meaningful and accurate measure of chlorophyll-a can be obtained. The Scale Factor is correct as long as the condition of the plankton population does not change; the condition does change with season and geographic location.

Example Calculation of Scale Factor from field calibration:

Seawater sample analysis indicates chlorophyll-a concentration is 50 µg/l when fluorometer reads 3.2 volts; measured signal for a sea water blank is 0.05 volts.

concentration (µg/l) = (Vsample - Vblank) * Scale Factor
50 = (3.2 - 0.05) * Scale Factor

Solving for Scale Factor:
Scale Factor = (50) / (3.2 - 0.05) = 15.87 µg/l

Enter the new Scale Factor in the configuration (.con) file in SEASAVE and/or SBE Data Processing.

 

Turbidity Meter Calibration Coefficients

The ECO-FL-NTU’s turbidity meter has a response that is linear over the measurement range provided: 0 to 25 NTU. The turbidity channel in the ECO-FL-NTU is not currently directly supported in Sea-Bird software. However, you can set up the turbidity channel as a User Polynomial in the configuration (.con) file, which allows you to define an equation to relate the sensor output voltage to calculated engineering units:

Value = a0 + (a1 * V) + (a2 * V2) + (a3 * V3)

Wet Labs defines turbidity as:

turbidity (NTU)  = (Vsample - Vblank) * Scale Factor

where:

      Vsample (volts) = in situ output of the turbidity meter

      Vblank (volts) = measured output for a seawater blank (pure, de-ionized water)

      Scale factor (NTU/volts) = multiplier

The ECO-AFL-NTU comes with a calibration sheet that lists values for Vblank and Scale Factor for the turbidity meter.

Setting the Wet Labs equation equal to the user polynomial equation and calculating a0, a1, a2, and a3:

(Vsample – Vblank) * scale factor = a0 + (a1 * V) + (a2 * V2) + (a3 * V3)

Expanding the left side of the equation and using consistent notation (Vsample = V),

scale factor * V – scale factor * Vblank = a0 + (a1 * V) + (a2 * V2) + (a3 * V3)

Rearranging:

(– scale factor * Vblank)  + (scale factor * V) = a0 + (a1 * V) + (a2 * V2) + (a3 * V3)

a0 = - scale factor * Vblank          a1 = scale factor          a2 = a3 = 0

Select the User Polynomial for the turbidity channel when setting up the configuration (.con) file in Sea-Bird software (SBE Data Processing or SEASAVE in SEASOFT-Win32 or SEACON in SEASOFT-DOS). The software prompts for a0, a1, a2, and a3.

While the factory-supplied Scale Factor can be used to obtain approximate values, field calibration is highly recommended. The relationship between turbidity and NTU is highly variable, and is not easy to determine in the laboratory. Particle shape and size are some of the factors that affect the relationship. To accurately measure NTU with a turbidity meter, perform calibrations on seawater samples with distributions of particles that are similar to what is expected in situ. Determine NTU independently, and use those values, as well as readings from the turbidity meter, to determine the correct Scale Factor. The Scale Factor is correct as long as the distribution of particle sizes and shapes does not change; the condition does change with season and geographic location.