Sea-Bird Electronics              Products            Support            Software            Sales            Service            Search/Sitemap

APPLICATION NOTE NO. 84

Using Instruments with Druck Pressure Sensors in Muddy or Biologically Productive Environments
July 2006

div.gif (3429 bytes)

Application note in pdf format

This Application Note applies to Sea-Bird instruments with Druck pressure sensors, for moored applications or other long deployments that meet either of the following conditions:

Note: Photo is for an SBE 37-SM. Pressure port details are similar for all instruments included in this application note.

At Sea-Bird, a pressure port plug with a small (0.042-inch diameter) vent hole in the center is inserted in the pressure sensor port. The vent hole allows hydrostatic pressure to be transmitted to the pressure sensor inside the instrument.
  • If the instrument is deployed in a high-sediment (muddy) environment with the pressure sensor end up, the pressure port may partially fill with sediment (through the vent hole) over time, causing a delay in the pressure response.
  • If the instrument is deployed in a biologically productive environment, the vent hole may be covered with biological growth over time, causing a delay in the pressure response, or in extreme cases completely blocking the pressure signal.
Sea-Bird has developed a high-head pressure port plug for deployment in muddy and/or biologically productive environments. The high-head plug extends beyond the surface of the instrument end cap, and has four horizontal vent holes connecting internally to a vertical vent hole.
  • The horizontal orientation of the external holes prevents the deposit of sediment inside the pressure port.
  • Each of the four vent holes is larger (0.062-inch vs. 0.042-inch diameter) than the single vent hole in the standard pressure port plug, significantly reducing the possibility that biological growth will cover all of the hole(s).

To purchase the high-head pressure port plug, Part Number 233186, contact Sea-Bird.

 

High-Head Pressure Port Plug Installation
  1. Unscrew the standard pressure port plug from the pressure port.
  2. Rinse the pressure port with warm, de-ionized water to remove any particles, debris, etc. Do not put a brush or any object in the pressure port; doing so may damage or break the pressure sensor.
  3. Install the high-head pressure port plug in the pressure port.

Note: Until several years ago, Sea-Bird filled the pressure port with silicon oil at the factory. For Druck pressure sensors, we determined that this was unnecessary, and no longer do so. It is not necessary to refill the oil in the field. However, for Paine or Paroscientific Digiquartz pressure sensors, the pressure port does need to be refilled with silicon oil. Please contact Sea-Bird with the serial number of your instrument if you are unsure of the type of pressure sensor installed in your instrument.

div.gif (3429 bytes)

Last updated: 06 Apr 2007

Sea-Bird Home     Phone: 425-643-9866     Fax: 425-643-9954     E-mail: seabird@seabird.com